Título : | Direct injection nebulizer with replaceable capillary for micro samples analysis by inductively coupled plasma-optical emission spectrometry |
Autor : | Maldonado, Domingo Chirinos, José Benzo, Zully Marcano, Eunice Gómez, Clara Salas, Janeth Salas1 Quintal, Manuelita D’Suze, Gina |
Palabras clave : | Replaceable capillary direct injection ICP-OES |
Fecha de publicación : | 8-Feb-2017 |
Resumen : | The analytical performance of a low cost
direct injection high efficiency nebulizer (DIHEN)
with replaceable capillary is evaluated and compared
with a high efficiency nebulizer HEN-cyclonic spray
chamber arrangement for sample introduction at lower
solution uptake rates in inductively coupled plasma
optical emission spectrometry (ICP-OES). The system
allows the use of a replaceable capillary and optimization
of the aerosol generation. The study was performed
in terms of the main analytical figure of merits
(i.e., sensitivity, limits of detection and signal stability)
for plain water solutions and matrix effects caused by
nitric acid and sodium in ICP-OES. In general, the
DIHEN with replaceable capillary offers better sensitivities
than the HEN-Cyclonic system and for some
elements, improved detection limits (DL) in the range
of 1.1–3.8 ng mL 1 for Ba,Mg, Zn 206.2–200 nm and
Zn 213.8–856 nm were obtained. Strontium’s DL was
comparable with both systems and those for Cd and
Mn were slightly worse. Short-term precision for this
unit is not as good as the HEN-Cyclonic system, but
being less than 2%, it can still be considered acceptable.
It was more susceptible to changes in matrix
composition, being ca. 18% the acid depressive effect
and 7% due to sodium with the DIHEN compared to
the conventional nebulizer-spray chamber arrangement
(13 and 3% acid and sodium depressive effect,
respectively). The utility of the developed unit is explored
analyzing three certified materials (bovine liver,
rice and wheat flours) and the experimental results are
within the certified values. The ability of the DIHEN
with replaceable capillary to analyze real micro samples
(samples reconstituted in 150 mL deionized water)
was demonstrated by the determination of Ca, Cd, Cu,
Mg, Mn, Sr, and Zn in individual scorpion venoms
samples. One of the main advantages over other
arrangements is that it can be easily constructed in
the laboratory. |