Título : | Mixed-gas inductively coupled plasma atomic emission spectrometry using a direct injection high efficiency nebulizer |
Autor : | Chirinos, José Kahen, Kaveh O´Brien, Su-Ann Montaser, Akbar |
Palabras clave : | Direct injection high efficiency nebulizer DIHEN Mixed-gas inductively coupled plasmas Atomic emission spectrometry Matrix interferences Computer simulation |
Fecha de publicación : | 2002 |
Editorial : | Analytical and Bionalytical Chemistry |
Citación : | Anal Bioanal Chem (2002) 372 :128–135 |
Resumen : | Experimental studies and computer simulations
were conducted to identify plasma operating conditions
and to explore and contrast the excitation conditions of
Ar, Ar-O2, and Ar-He inductively coupled plasmas (ICPs)
for the introduction of microliter volumes of sample solutions
with a direct injection high efficiency nebulizer
(DIHEN). The best MgII 280.270 nm/MgI 285.213 nm ratio
(6.6) measured with Ar ICP atomic emission spectrometry
for the DIHEN (RF power = 1500 W; nebulizer gas
flow rate = 0.12 L min–1) was less than the ratio (8.2) acquired
on the same instrument for conventional nebulization
(1500 W and 0.6 L min–1). Addition of small amounts
of O2 or He (5%) to the outer gas flow improved excitation
conditions in the ICP, that is, a more robust condition
(a MgII/MgI ratio of up to 8.9) could be obtained by using
the DIHEN with Ar-O2 and Ar-He mixed-gas plasmas,
thereby minimizing some potential spectroscopic and matrix
interferences, in comparison to Ar ICPAES. |