SABER UCV >
1) Investigación >
Artículos Publicados >
Por favor, use este identificador para citar o enlazar este ítem:
http://hdl.handle.net/10872/4353
|
Título : | Simultaneous determination of Cr, Ni and V in urine by ET AAS |
Autor : | Hermida, Jeimy Alvarez, María A. |
Palabras clave : | simultaneous analysis; ET AAS; metals in urine |
Fecha de publicación : | 2011 |
Editorial : | Sociedade Brasileira de Química, http://quimicanova.sbq.org.br/index.php |
Citación : | Química Nova, 2011 (34) 940-944 |
Resumen : | SIMULTANEOUS DETERMINATION OF Cr, Ni AND V IN URINE BY ET AAS. A procedure for the simultaneous determination
of Cr, Ni, and V in urine by electrothermal atomic absorption spectrometry (ET AAS) was optimized by factorial design, and
performed at a pyrolysis and atomization temperatures of 1300 and 2500 ºC, respectively, using 15 μg de Mg(NO3)2 as chemical
modifier. Characteristics mass of 14, 6 and 220 rg and detection limits of the method of 0.07, 0.38 and 0.75 µg L-1 were obtained for
Cr, Ni and V respectively. The methodology was validated using a Liphochek Urine Metals Control sample (Bio-Rad) (P=0.05). The
methodology was applied to samples of voluntary Venezuelan people, not environmentally exposed to specific emissions, and results
ranging from < LOD-1.1 and 1.3-3.3 µg L-1 was observed for Cr and V, respectively, and not detectable levels for Ni. |