Título : | Liquid sampling-atmospheric pressure glow discharge as a secondary excitation source: Assessment of plasma characteristics |
Autor : | Manard, Benjamin T. Gonzalez, Jhanis J. Sarkar, Arnab Dong, Meirong Chirinos, José Mao, Xianglei Russo, Richard E. Kenneth Marcus, R. |
Palabras clave : | LS-APGD Laser ablation Liquid sampling-atmospheric pressure glow discharge |
Fecha de publicación : | 3-Jun-2015 |
Resumen : | The liquid sampling-atmospheric pressure glow discharge (LS-APGD) has been assessed as a secondary excitation
source with a parametric evaluation regarding carrier gas flow rate, applied current, and electrode distance.
With this parametric evaluation, plasmaopticalemission was monitored in order to obtain a fundamental understanding
with regards to rotational temperature (Trot), excitation temperature (Texc), electron number density
(ne), and plasma robustness. Incentive for these studies is not only for a greater overall fundamental knowledge
of the APGD, but also in instrumenting a secondary excitation/ionization source following laser ablation (LA). Rotational
temperatures were determined through experimentally fitting of the N2 and OH molecular emission
bandswhile atomic excitation temperatureswere calculated using a Boltzmann distribution of He andMgatomic
lines. The rotational and excitation temperatureswere determined to be ~1000 K and ~2700 K respectively. Electron
number density was calculated to be on the order of ~3 × 1015 cm−3 utilizing Stark broadening effects of the
Hαline of the Balmer series and a He I transition. In addition, those diagnosticswere performed introducing magnesium(
by solution feed and laser ablation) into the plasma in order to determine any perturbation under heavy
matrix sampling. The so-called plasma robustness factor, derived bymonitoringMg II/MgI emission ratios, is also
employed as a reflection of potential perturbations in microplasma energetics across the various operation conditions
and sample loadings. While truly a miniaturized source (b1 mm3 volume), the LS-APGD is shown to be
quite robust with plasma characteristics and temperatures being unaffected upon introduction of metal species,
whether by liquid or laser ablation sample introduction. |